40 (67) BRUKER BVT3000 Version 003
Remote interface control
NH (Nitrogen heater) 4.8.4
Write
Syntax: EOT0000STXNH Value ETX BCC
Response
:ACK
Description
: This command allows the liquid nitrogen heater power (Evaporator)
to be controlled.
Rules:
The value from 0 up to 100%, is defined as a string up to 5 charac-
ters long. The string can begin with 1 to 5 spaces or «0». After power on, the initial
valueissetto0(thenitrogenheaterisOFF).
Read
Syntax: EOT0000NH ENQ
Response
:STXNH Value ETX BCC
Description
: This command allows the liquid nitrogen heater power to be read
back.
Rules
: Value from 0 to 100%, is a string up to 5 characters long. The string
can begin with 1 to 5 spaces or «0».
NB: Value is a DECIMAL code.