T104 Series Indicators EN-23
2. The first digit is displayed flashing.
3. Press the No button to increase the digit or press the
Yes button to accept the digit and move to the next
one.
Note: press the Back button can decrease the digit.
4. Repeat this process for all digits.
5. Press the Yes button when the last digit has been set.
6. The new setting is displayed with all digits flashing.
Press the Yes button to accept the setting or press the
No button to resume editing.
7. To end the current menu selection, press the Yes
button to advance to the next menu, or press the No
button to return to the top of the current menu.
4.2 Calibration Menu
Enter the calibration menu to perform calibrations.
4.2.1 Initial Calibration
When the scale is operated for the first time, span calibration is recommended to ensure accurate weighing
results.
Before performing the calibration, be sure to have the appropriate calibration weights as listed in table 4-1.
Ensure that the LFT switch/calibration lock is set to the unlocked position.
You can also adjust the GEO setting according to your location
.
For detailed GEO information, please check
the Geo Code Adjustment section below.
Table 4-1 Suggested Span Calibration Mass (Sold Separately)
Max Capacity (kg) Mass
*
(kg) Max Capacity (lb) Mass* (lb)
3 kg 3 kg 6 lb 6 lb
6 kg 6 kg 12 lb 12 lb
15 kg 15 kg 30 lb 30 lb
30 kg 30 kg 60 lb 60 lb
60 kg 60 kg 120 lb 120 lb
150 kg 150 kg 300 lb 300 lb
Note:
The calibration unit is consistent with the capacity unit.