EN-22 T104 Series Indicators
4. MENU SETTINGS
The User Menu allows the customizing of scale settings.
Note: Sub-menu for options (USB, Ethernet and Print2 in the table below) will be active only when the specific
board is installed. See their separate option instruction manuals for additional setting information.
4.1 Menu Navigation
4.1.1 User Menu (in segments)
Notes:
Some modes/units may not be available in all models.
When LEGAL FOR TRADE is turned ON (the security switch S1 is in the locked position), the menu settings
will be affected as below:
Calibration function is disabled.
Zero Range setting is locked at 2%.
Stable Range setting is locked at 1d.
Auto-Zero Tracking setting is locked at 0.5d.
Filter and Units are locked at their current settings.
Stable Only is locked at On.
Lb
:
oz unit is locked at Off.
Power saving mode is disabled.
Note: Please refer to Legal for Trade chapter for the position of the security switch S1.
4.1.2 Button Navigation
The Yes button allows entry into the displayed menu.
Accepts the displayed setting and advances to the next item.
The No button rejects entry into the displayed menu.
Rejects the displayed menu and moves on to the next selection.
The Back button: moves backwards to the previous menu.
Returns to the previous displayed menu or selection.
The Exit button: exits from user menu to the active weighing mode.
For menu items which need to set numbers such as the scale capacity, the current
number is displayed with all digits flashing. To revise:
1. Press the No button to begin editing.