T104 Series Indicators EN-27
Grad2
[ ]
Set the second readability of the scale.
Note:
The value of graduation 2 must be larger than graduation 1, and it applies to capacity as well.
The capacity and graduation value should be within the following range, or your setting will be
unsuccessful:
Power On Zero
[ ]
Zero the scale at Power On.
= disabled.
= enabled.
Power On Unit
[ ]
Set the unit that will be displayed at Power On.
= last unit in use when turned off
kg = kilograms
g = grams
Ib = pounds
oz = ounces
Ib:oz = pound ounces
t = metric tonne
Auto Tare
[ ]
Set the automatic tare feature.
= automatic tare is disabled.
= the first stable gross weight is tared. The tared value will be cleared when the indicator returns
to gross zero.
= stable gross weights within the Accept limits are tared (in Check mode only).
4.4 Readout Menu
Enter this menu to set user preferences. Default settings are in bold.
Menu Sub-Menu
Sub-Menu (in
segment)
Options Options ( in segment)
Read Out
( )
Reset
no, yes
,
Stability
0.5d, 1d, 2d, 5d
, , ,
Zero Range
2%, 100%
,
Filter Level
Low, Medium, High
, ,
Auto Zero Track
Off, 0.5d, 1d, 3d
, , ,
Back Light
Off, On, Auto
, ,
Back Light Color
Green, Yellow, Red,
Amber
, , ,
Screen Saver
Off, 1min, 2min, 5min
, , ,
Auto Off Off, 1min, 5min, 10min
, , ,
PWM.SAVE
ON, OFF
,
CEC ON, OFF
,
End
/ /
Note: / refers to not applicable in the table above.
Reset [ ]
Reset the readout menu to factory defaults.
= do not reset.
= reset