BÜCHI Labortechnik AG Installation | 5
Operation Manual ProxiMate™ 25/94
Navigation path
➔ ➔ [Calibrate BCV]
Precondition:
R The instrument is in administrator mode. See Chapter6.5 "Log in administrator
mode", page39
R A sample presentation is defined.
R The instrument is stabilized with at least 2 hours continuous running.
u Navigate to the action [Calibrate BCV] via the navigation path.
ð The instrument shows a dialog box with mandatory settings.
Fig.10: Dialog box Calibrate BCV
1 Measurement view 2 External Reference
3 Progress (view) 4 Measurement mode
5 Sample presentation ID
u Select the name of the sample presentation you wish to calibrate.
u Enter the corresponding values.
u Place the white reference or empty sample presentation method in the correct po-
sition.
u Tap the button [Start].
u Follow instructions on the display during the calibration.
u Confirm the instructions by tapping the [OK] button.
5.6.4 Import external reference data
NOTE
The location of the destination folder is fixed. See Chapter10.2 "Folder locations",
page89
Navigation path
➔ ➔ [Import External Refer-
ence Data]
Precondition:
R The reference data file is saved on the instrument.
u Navigate to the action [Import External Reference Data] via the navigation path.
ð The display shows the dialog box Import External Reference Data.