6.9 Deleting a property
NOTE
It is not possible to delete a property after it has been used for measurement.
Proceedings:
Navigation path
➔
Precondition:
R The instrument is in administrator mode. See Chapter6.5 "Log in administrator
mode", page39
u Navigate to the [Application] menu via the navigation path.
u Tap the function [Edit] on the function bar.
ð The display highlights the function [Edit] in green.
u Tap the name of the application that you wish to edit.
ð The display highlights the application in green.
u Tap the name of the property you wish to remove.
ð The display highlights the application in green.
u Tap the function [Remove] on the function bar.
u Select [Yes] to confirm the action in response to the confirmation question.
ð The property is deleted.
6.10 Adding measurement data to a calibration
6.10.1 Performing a calibration update with AutoCal
For performing the AutoCal function the following measurement data is needed:
minimum three measurements
Two of these three measurements with different reference values
Navigation path
➔
Precondition:
R The instrument is in administrator mode. See Chapter6.5 "Log in administrator
mode", page39
R Reference values are set. See Chapter6.10.2 "Setting a reference value to a
property", page66
R The measurement you wish to use is activated for AutoCal. See Chapter6.10.3
"Activating a measurement for AutoCal", page66
u Navigate to the [History] menu via the navigation path.
u Tap the function [Edit] on the function bar.
ð The display highlights the function [Edit] in green.
u Tap the name of the measurement that you wish to use.
u Tap the button [ AutoCal] on the function bar.
ð The display shows the dialog AutoCal.