BÜCHI Labortechnik AG Operation | 6
Operation Manual ProxiMate™ 45/94
Proceedings:
Navigation path
➔
Precondition:
R The instrument is in administrator mode. See Chapter6.5 "Log in administrator
mode", page39
u Navigate to the [Application] menu via the navigation path.
u Tap the function [Edit] on the function bar.
ð The display highlights the function [Edit] in green.
u Tap the name of the application that you wish to edit.
ð The display highlights the application in green.
u Tap the action [Rotation].
ð The display shows a dialog box with the selectable values.
u Select a value.
u Tap the button [OK].
ð The dialog box closes.
ð The setting is saved.
6.6.9 Changing the Measurement duration of an application
The default measurement time (15 seconds) allows the sample to complete one rota-
tion. Shorter measurement time does not allow a complete rotation of the sample.
Proceedings:
Navigation path
➔
Precondition:
R The instrument is in administrator mode. See Chapter6.5 "Log in administrator
mode", page39
u Navigate to the [Application] menu via the navigation path.
u Tap the function [Edit] on the function bar.
ð The display highlights the function [Edit] in green.
u Tap the name of the application that you wish to edit.
ð The display highlights the application in green.
u Tap the action [Measurement duration].
ð The display shows a dialog box with a numeric input box.
u Enter the value in the numeric input box.
u Tap the button [OK].
ð The dialog box closes.
ð The value for the duration is saved.
6.6.10 Enter a Standard Operating Procedure (SOP) for an application
Each Application has an optional Standard Operating Procedure (SOP) associated
with it. A Standard Operating Procedure SOP shows instructions for the user to follow
during measurement.