BÜCHI Labortechnik AG Operation | 6
Operation Manual ProxiMate™ 63/94
Precondition:
R The instrument is in administrator mode. See Chapter6.5 "Log in administrator
mode", page39
u Navigate to the [Application] menu via the navigation path.
u Tap the function [Edit] on the function bar.
ð The display highlights the function [Edit] in green.
u Tap the name of the application that you wish to edit.
ð The display highlights the application in green.
u Tap the name of the property that you wish to edit.
ð The display shows the view Property.
u Tap the action [Limit Max.].
ð The control panel shows a dialog box with a numeric input box.
u Enter the value in the numeric input box.
u Tap the button [OK].
ð The dialog box closes.
ð The value for the Limit is saved.
6.8.22 Changing the Limit minimum of a property
Sets the lower limit for the property.
Proceedings:
Navigation path
➔
Precondition:
R The instrument is in administrator mode. See Chapter6.5 "Log in administrator
mode", page39
u Navigate to the [Application] menu via the navigation path.
u Tap the function [Edit] on the function bar.
ð The display highlights the function [Edit] in green.
u Tap the name of the application that you wish to edit.
ð The display highlights the application in green.
u Tap the name of the property that you wish to edit.
ð The display shows the view Property.
u Tap the action [Limit Min.].
ð The control panel shows a dialog box with a numeric input box.
u Enter the value in the numeric input box.
u Tap the button [OK].
ð The dialog box closes.
ð The value for the Limit is saved.
6.8.23 Changing the Tolerance maximum of a property
Tolerance is a difference from the target value.
Sets the upper tolerance relative to the target.
Proceedings: