7 | Cleaning and servicing BÜCHI Labortechnik AG
84/94 Operation Manual ProxiMate™
u Navigate to the action [Test BCV] via the navigation path.
ð The display shows the dialog Test BCV.
u Select the name of the sample presentation you wish to test.
u Tap the button [Start].
u Follow instructions on the display during the test.
u Confirm the instructions by tapping the [OK] button.
u Once the test is complete, the instrument records the results in a report.
ð The display shows Baseline Correction Tests is completed.
7.5.2 Carrying out a Comprehensive System Test
Navigation path
➔ ➔ [Extended System Test]
Precondition:
R The instrument is in administrator mode. See Chapter6.5 "Log in administrator
mode", page39
R External reference data for all standards are loaded.
R Performance Test Standards Kit is available.
u Navigate to the action [Extended System Test] via the navigation path.
ð The display shows the dialog Extended System Test.
u Select check box [Comprehensive System Test]
u Tap the button [Start].
u Follow instructions on the display during the test.
u Confirm the instructions by tapping the [OK] button.
u Once the test is complete, the instrument records the results in a report.
ð The display shows Comprehensive System Tests is completed.
7.5.3 Carrying out an Advanced System Test
Navigation path
➔ ➔ [Extended System Test]
Precondition:
R The instrument is in administrator mode. See Chapter6.5 "Log in administrator
mode", page39
u Navigate to the action [Extended System Test] via the navigation path.
ð The display shows the dialog Extended System Test.
u Select check box [Advanced System Test]
u Tap the button [Start].
ð The instrument starts the test.
u Once the test is complete, the instrument records the results in a report.
ð The display shows Advanced System Tests is completed.
7.6 Carrying out a data backup
NOTE
The location of the destination folder is fixed. See Chapter10.2 "Folder locations",
page89