BÜCHI Labortechnik AG Operation | 6
Operation Manual ProxiMate™ 55/94
6.8.8 Changing the Calibration Base (Calibration model only)
The following settings are available:
Type Explanation
[Dry Basis] The property value is expressed neglecting the presence
of water in the sample.
The water contribution is subtracted from the calculation.
[Moisture xx %] The property value is expressed assuming xx % water
content.
The amount of water is expressed as a percentage of the
total weight.
[As Is] The property value is expressed including the presence of
water in the sample.
The water contribution is included in the calculation.
Proceedings:
Navigation path
➔
Precondition:
R The instrument is in administrator mode. See Chapter6.5 "Log in administrator
mode", page39
R The post prediction type [Calibration Base Standard] is selected.
u Navigate to the [Application] menu via the navigation path.
u Tap the function [Edit] on the function bar.
ð The display highlights the function [Edit] in green.
u Tap the name of the application that you wish to edit.
ð The display highlights the application in green.
u Tap the name of the property that you wish to edit.
ð The display shows the view Property.
u Tap the action [Calibration Base].
ð The display shows a dialog box with the selectable values.
u Select a value.
u Tap the button [OK].
ð The dialog box closes.
ð The setting is saved.
ð The value for the calibration base is saved.
6.8.9 Changing the Display Basis (Calibration model only)
The following settings are available:
Type Explanation
[Dry Basis] The property value is expressed neglecting the presence
of water in the sample.
The water contribution is subtracted from the calculation.