BÜCHI Labortechnik AG Cleaning and servicing | 7
Operation Manual ProxiMate™ 85/94
Navigation path
➔ ➔ [Backup NIRWise Data]
Precondition:
R The instrument is in administrator mode. See Chapter6.5 "Log in administrator
mode", page39
u Navigate to the action [Backup NIRWise Data] via the navigation path.
ð The display shows a dialog with values that can be saved.
u Select the values according to your needs.
u Tap the button [Start].
ð The instrument creates a .zip file with the selected data.
u Save the data to an external data store.
7.7 Cleaning the display
The cleaning mode switches off the touch function of the touchscreen.
Navigation path
➔ ➔ [Cleaning Mode]
u Navigate to the action [Cleaning Mode] via the navigation path.
ð The touch function of the display is locked.
ð The display shows a dialog box with a numeric input box.
u Wipe down the display with a damp cloth.
u Enter the displayed secure code.
u Tap the button [OK].
ð The dialog box closes.
ð The touch function of the display is unlocked.