7 | Cleaning and servicing BÜCHI Labortechnik AG
80/94 Operation Manual ProxiMate™
7.2.3 Confirm Lamp Replacement
Navigation path
[Confirm Lamp Replace-
ment]
Precondition:
R The lamp has been changed.
R The instrument is in administrator mode. See Chapter6.5 "Log in administrator
mode", page39
u Tap the On/Off master switch.
ð The system starts up.
ð The display highlights the status bar yellow.
ð After completion the startup phase the display highlights the status bar black.
u Navigate to the action [Confirm Lamp Replacement] via the navigation path.
u Select which lamp was replaced (up view lamp or down view lamp).
u Confirm the secure question with OK.
7.3 Checking the desiccant cartridge
Color Describtion
Pink used cartridge (replacement required)
Blue unused cartridge
u Open the cover lid for the desiccant
cartridge.
u Remove the cover lid and the seal.