Why is the displayed flow rate always nil on my Bürkert Measuring Instruments?
- Jjeffrey63Aug 30, 2025
If the displayed flow rate on your Bürkert Measuring Instruments is always nil, ensure the flow rate in the pipe is not nil, check that the entered K factor is correct. Also, consider choosing a smaller flow rate unit or increasing the number of displayed decimals.