6.2.2 Inspection report
Unplug the switch mode power supply from the electrical socket.
Unplug the 24-volt power supply cable from the socket.
Insert the plug of the 24-volt cable into the measuring adapter (special accessory,
available from BURMEIER on request).
Connect the measuring adapter to the “test probe” or similar socket of the test de-
vice.
Plug the switch mode power supply into the test socket on the test device.
If equipped with a CA40 control unit or an external switch mode power supply:
Plug the mains plug/switch mode power supply of the bed into the test socket of
the measuring device.
Connect the test device probe to a bare metal conductive part of the bed frame
(e.g. a screw).
Test procedure:
Leakage current test: direct or differential current in accordance with DIN EN 62353
Perform a leakage current test in accordance with the instructions provided by the test
device manufacturer.
Limit value:
Leakage current IL must be less than 0.1 mA.
Inspection cycle:
We recommend an annual inspection and functional check. If this test has been passed, an
electrical measurement every ten years is sufficient if the bed is equipped with an external
switch mode power supply. When equipped with a CA 40 control unit, the electrical measure-
ment must be carried out annually as a regular part of the inspection and function check. In
the case of verifiable compliance with 2% error rate (see also DGUV regulation 3: § 5, table
1B), the inspection cycle of the electrical test and inspection and function test can be exten-
ded to a maximum of 2 years.
The inspection report templates shown on the pages that follow should be used.
The following is an inspection report template for inspecting electro-medical equipment in ac-
cordance with DIN EN 62353 (latest issue):
Inspection report