11.1.2 Safety Side Requirements
11.2 Translation of EC Declaration of Conformity
Safety Side Requirements
Height above mattress > 220 mm
Gap between bars and mattress base < 120 mm
Foam rubber density min. 38 kg/m
3
Gap between safety side and chassis head-
board
< 60 mm
Gap between safety side and chassis footboard >318 mm
Permissible accessory safety sides:
Type/model “ASG” adjustable safety sides
Item no.: (right-hand side) 250198
Item no.: (left-hand side) 250139
We, Burmeier GmbH & Co. KG, in our sole responsibility as the manufacturer, hereby de-
clare that this product complies with the provisions of REGULATION (EU) 2017/45 OF THE
EUROPEAN PARLIAMENT AND THE COUNCIL of 5 April 2017 (MDR).
The full latest version of the declaration of conformity is available on request from our cus-
tomer centre (for contact details please refer to the chapter Part A: Address, information for
customers, market note » 1) or go to the dealer area on our website.
Appendix
Part B: Operator and Technical Personnel 51