PRELIMINARY
Document type: Title: Revision date: Revision:
User's Manual (MUT) N1471 4 Channel Programmable HV Power Supply 24/03/2010 1
NPO: Filename: Number of pages: Page:
00112/07:N1471.MUTx/01 N1471_REV1.DOC 38 36
3.5.4.1. Meaning of Board Alarm bits
Bit Function
Bit 0 → CH0 1 : Ch0 in Alarm status
Bit 1 → CH1 1 : Ch1 in Alarm status
Bit 2 → CH2 1 : Ch2 in Alarm status
Bit 3 → CH3 1 : Ch3 in Alarm status
Bit 4 → PWFAIL 1 : Board in POWER FAIL
Bit 5 → OVP 1 : Board in OVER POWER
Bit 6 → HVCKFAIL
1 : Internal HV Clock FAIL (≠ 200±10kHz)
3.5.5. SET commands related to the Channels
The following table contains the strings to be used to handle set commands related to the
Channels.
The 'X' in the Field 'Channel' can be set to the '0..4' values.
When 'X=4' the command is issued to all 4 Channels.
String Function (Units)
$BD:xx,CMD:SET,CH:X,PAR:VSET,VAL:XXXX.X Set VSET value
$BD:xx,CMD:SET,CH:X,PAR:ISET,VAL:XXXX.XX Set ISET value
$BD:xx,CMD:SET,CH:X,PAR:MAXV,VAL:XXXX Set MAXVSET value
$BD:xx,CMD:SET,CH:X,PAR:RUP,VAL:XXX Set RAMP UP value
$BD:xx,CMD:SET,CH:X,PAR:RDWN,VAL:XXX Set RAMP DOWN value
$BD:xx,CMD:SET,CH:X,PAR:TRIP,VAL:XXXX.X Set TRIP time value
$BD:xx,CMD:SET,CH:X,PAR:PDWN,VAL:RAMP/KILL Set POWER DOWN mode value
$BD:xx,CMD:SET,CH:X,PAR:IMRANGE,VAL:HIGH/LOW Set IMON RANGE value
7
$BD:xx,CMD:SET,CH:X,PAR:ON Set Ch ON
$BD:xx,CMD:SET,CH:X,PAR:OFF Set Ch OFF
3.5.6. SET commands related to the module
String Function (Units)
$BD:xx,CMD:SET,PAR:BDILKM,VAL:OPEN/CLOSED Set Interlock Mode
$BD:xx,CMD:SET,PAR:BDCLR Clear alarm signal
7
parameter 'IMRANGE' can be changed only on modules featuring IMON zoom (optional)