DRSplus Operating Manual
13/58
drsplus needs to be operated in the following environmental conditions:
o Temperature: +10 °C to +35 °C
o Humidity (max): 90% not condensing
drsplus needs to be stored in the following environmental conditions:
o Temperature: -10 °C to +60 °C
o Humidity (max): 90% not condensing
The device must be placed in a room which is not exposed to adverse chemical-
physical conditions, such as the presence of sulfur, salt, dust, direct sunlight, lack
of ventilation, high humidity, sudden temperature drops or peaks. The safety
and/or effectiveness of the instrument cannot be guaranteed if these conditions
are not met.
The following precautions are important to avoid incorrect use of the device:
Provide explanations to patients before placing them in front of the device.
Use the device in dim light, or at least away from direct light. This will facilitate the
natural dilation of the pupil.
The minimum pupil diameter required to obtain good quality images is 2.5 mm.
If the patient does not fix correctly and steadily, the images acquired may relate
to portions of retina that are not what is expected.