DRSplus ā Operating Manual
16. Cleaning
This paragraph explains how to clean the device. The device must be powered off, and the power cord
shall be disconnected from mains.
The front lens should be cleaned using a small hand pump air blower to blow away dust.
If the lens is very dirty, for instance due to the presence of fingerprints or other impurities, the front
lens should be cleaned using photographic cleaning paper or a very clean microfiber cloth and a
suitable lens cleaning fluid.
Pass a wet wipe on the front lens with a single circular motion: never reuse it after each pass. Several
passes may be needed in order to achieve a good cleaning level.
Do not attempt to clean the front lens with a dry cloth as this may scratch the surface.
The headrest silicon cushion is the only part in direct contact with the patient: it should be sanitized
with a disinfecting wipe after each use and allowed to dry prior to reuse.
Take care not to sprinkle parts not belonging to the patient rest. The headrest silicon cushion can also
be removed and washed with lukewarm water and a mild detergent.
The touch screen panel should be cleaned using a soft, lint-free cloth dampened with a small amount
of water.
Do not use alcohol or detergents to clean the touch screen, as these may damage the
protective coating.
The plastic covers of the device can be cleaned using of a cloth dampened in a small amount of water.
Do not use alcohol or detergents to clean the plastic shells, as this could erase
labelling and other indications.