12.3.4.2 Checking for High Power Range
A. Press CP and RANGE to select CPH. The screen shows:
CPHA: 125.00W
CPHB: 125.00W
B. The current shunt range is 250A. Input the power listed in Table 12-11. Press LOAD
ON/OFF to start the Load. Refer to the DMM (V) value and adjust the DC Power
Supply value to be the same as the previous test model, and then wait for 30 seconds.
Record the DMM (V) voltage passed through the Load input terminal and the DMM (I)
reading of shunt current.
Table 12-11
Model
Front Panel Reading
63201
63202_500V
63202_600V
63203
63204_500V
63204_600V