General Information
37
4. General Information
4.1 Tester Maintenance
The Easy-Touch Pro tester requires no maintenance. If desired you may clean the outside surfaces of the tester. Since some
cleaning agents leave a conductive residue, take special care to not allow the cleaning agents to come in contact with the
test point connectors or the circuitry inside the casework.
4.2 Fixture Maintenance
The xturing contacts that mate to the device-under-test may wear due to repeated insertion cycles. Contact wear can
result in higher connection resistances which increases the measured resistances for the device-under-test. For this reason
Cirris recommends that customers evaluate the number of mating cycles and the cycle life of xture contacts to determine
maintenance intervals for testing and/or replacing xturing contacts. A good way to check xture contact resistance is to
construct and use a shorting block. For more information on creating shorting blocks, see http://www.cirris.com/adapters/
test-adapt.html.
4.3 Service
Cirris Testers are modular in design for easy servicing. If your Cirris tester requires service, as determined by Cirris support
personnel, you may need to send the aected module or the entire tester back to Cirris for repair. A loaner tester can be
sent to you. You should not attempt to service any circuit board at the component level. All component-level service should
be performed by Cirris technicians.
4.4 Calibration
With your Easy-Touch Pro tester you should have received a Certicate of Verication. Before leaving the factory, every
Easy-Touch Pro tester is calibrated in compliance with ANSI/NCSL Z540-1-1994 and MIL-STD-45662A to standards traceable
to the NIST in the United States. The calibration of the tester should thereafter be veried annually. To verify calibration and
functionality, you can purchase the Easy-Touch Pro Performance Check Kit. In addition to the performance check kit, you
need a calibrated volt meter and high voltage probe capable of measuring up to 1500 VDC.
Note: In the event a Cirris tester is found to be out of calibration, there are no adjustable controls. The tester, or aected
portions of the tester, must be sent back to a Cirris facility for repair.
4.5 Tester Symbols
Caution: See manual for more information Risk of shock