For better access to the screws of the power sources, it is recommended to remove the hydraulic
deposit tank to facilitate the operation. Install the new power supply and reconnect the cables.
Proceed starting a treatment and measure the source output voltage.
3.2.8 - Equipment does not decrease below 0 ° C
When configuring a treatment the temperatures can not be lowered below 0°C, it is because during
equipment update either the EEPROM was not programmed or or was programmed with an unused
list.
To check this access the information screen and compare the list number.