Remove the covers to access the
connector
Pin 4 and 5 of the lower connector
3.1.7 - Applicator temperature error
Problem in some temperature sensor or difference in temperature between the plates of a hand
piece. The treatment will be cancelled in case of having two faulty sensors in the same applicator, or
continue working if it is only one.
Whichever of this errors appears, it is recommended to replace the handpiece, although it is still
working.
There are 5 possible temperature reasons for this alarm, to find it out, it is necessary to access the
code screen, enter 123789 and press enter. The last 6 temperature alarms will appear, starting with
the newest (right) and the oldest (left). Each number has the following meaning:
0 : No alarm
1 : NTC sensor short circuit
2 : NTC sensor open circuit