incorrect damage or the HR channel is
wrong
The electrodes are connected
poorly
Reconnect the electrode
The wire between ECG
module and main board
poorly connected
Reconnect the cable or replace it
ECG module defective Replace the ECG module
Motherboard defective Replace the Mother board
4.6.5 SpO
2
Failures
Symptoms Possible Cause Corrective Action
The SpO
2
no
waveform
and value
The SPO2 probe is defective Replace the SPO2 probe
The cable from the SPO2 module to
side plate socket is not connect well
Check the cable from SPO2 module to side plate
socket and reconnect them
Cables from Motherboard to SPO2
module are defective or poorly
connected.
1. Check that the cable between the Motherboard
to the SPO2 module is correctly connected.
2. Check that the cables and connectors are not
damaged
SPO2 module is defective Replace the SPO2 module
Motherboard is defective Replace the Motherboard
SpO
2
value
is incorrect
The probe type of the patient is wrong
Change a right probe type of patient
Long period of NIBP measurement
on the same limb or some other
condition such as shock or low temp
Start the SPO2 measurement in a normal state
SPO2 module is defective Replace the SPO2 module
Motherboard is defective Replace the Motherboard
4.6.6 NIBP Failures