33
6.6 TEMP Test
Tool required:
Resistance box (with accuracy above 0.1Ω)
Follow this procedure to perform the test:
a) Connect the two pins of any Temp connector of a module to the two ends of the resistance box using 2
wires.
b) Set the resistance box to 1354.9Ω (corresponding temperature is 37ºC).
c) Verify each Temp channel of the monitor and make sure that the displayed value is within 37 ±
0.1ºC.You can also use a patient simulator to perform the Temp test.