21
Cant start air
inflation
Cables from mother board to
NIBP module are defective or
poorly connected.
First, check that the cable between the motherboard to
the NIBP module is correctly connected. Then, check
that the cables and connectors are not damage.
The keyboard defective Replace the Keyboard
The wire from the NIBP module
to air pump is not connected well
Check the wire from NIBP module to air pump,
replace it if it’s broken
NIBP module is defective Replace the NIBP module
Mother board is defective Replace the mother board
Inflate air
repeatedly
(NO NIBP
values )
The NIBP settings are wrong
Check out the type of the patient in the patient
management
The accessories have air leakage
Check out the cuff ,the cuff pipe ,or the interface
socket
The NIBP pump has air leakage
Inflatable with numerical display, but the value is not
stable, it can be seen that the numeric declines
2in 1 functions board Replace the 2 in 1 functions board
The NIBP
values are
incorrect
The type of the patient is wrong
Check out the type of the patient in the patient
management
A continue NIBP measurement
for a Long time
The measurement should start every 5 minutes at least
The NIBP accessories have
leakage
1.The cuff is tied up too tight or too loose
2.The cuff has leakage or is defective
3. The NIBP tubing has leakage
The NIBP module is defective Replace the NIBP module
The Mother board is defective Replace the Mother board
4.6.7 RESP Failures
Symptoms Possible Cause Corrective Action
NO RESP
data
The RESP is disabled Enable the RESP on setup menu
ECG cable or electrode is
defected
Replace the ECG cable
Cables from Motherboard to
ECG module defective or poorly
connected.
1. Check that the cable between the Motherboard to the
ECG module is correctly connected.
2. Check that the cables and connectors are not damage
The cable from the ECG module Check the cable from ECG module to side plate socket