22
to side plate socket is not
connected well
and reconnect them
ECG module is defective Replace the ECG module
Motherboard is defective Replace the Motherboard
RESP wave
and reading
are
incorrect
The signal of patient is so weak
Check out the patient condition
The electrode is defective Replace the ECG probe
ECG module is defective Replace the ECG module
Motherboard is defective Replace the Motherboard
4.6.8 IBP Module Defective
Symptoms Possible Cause Corrective Action
The IBP has
no readings
and no
waveforms
The monitor is not equipped
with IBP module
Check out whether there is a IBP module in the monitor
IBP module is disabled
Please enable the IBP module, For more details, you can
refer to the user manual.
Cables from Motherboard to
IBP module are defective or
poorly connected
1. Check that the cable between the Motherboard to the
IBP module is correctly connected.
2. Check that the cables and connectors are not damage.
The cable from the IBP module
to side plate socket is not
connect well
Check the cable from IBP module to side plate socket
and reconnect them
IBP module is defective Replace the IBP module
Motherboard is defective Replace the Motherboard
IBP readings
are incorrect
The IBP settings are incorrect
1. Zeroing must be done before use
2. Disposable IBP sensor must be change after single
use
3. Ensure the channel you set is the channel you are
using
The IBP extension cable or
sensor is defected
Replace the extension cable or sensor
The IBP module is defective Replace the IBP module