23
4.6.9 TEMP Module Defective
Symptoms Possible Cause Corrective Action
NO TEMP values
The temperature is lower
than the monitor can
measure range
Check out the TEMP is in the normal range
The TEMP probe is
defective
Change another TEMP probe
Cables from Motherboard to
ECG module are defective or
poorly connected
1. Check that the cable between the Motherboard to
the ECG module is correctly connected.
2. Check that the cables and probe are not damage
The cable from the ECG
module to side plate socket
is not connected well
Check the cable from ECG module to side plate
socket and reconnect them
TEMP module is defective Replace the TEMP module
Motherboard is defective Replace the Motherboard
TEMP value is
incorrect
The type of temp sensor is
incorrect
it can supports CF and YSI, change the setting you
need
TEMP probe is incorrect Replace the TEMP probe
ECG module is defective Replace the ECG module
TEMP Value
unstable
The TEMP socket defective Check out whether it has a bad contact
The TEMP probe defective Change a TEMP probe
The ECG module is
defective
Replace the ECG module
The Mother board is
defective
Replace the Mother board
4.6.10 EtCO
2
Failures
Symptoms Possible Cause Corrective Action
NO EtCO2
readings and wave
Accessories (sampling line
or adapter) is defective
Replace sampling line or airway adapter.