34
GenesisPlus Operator Manual
D1151, R
EV C, 10/16
Inspect and Clean the Handpiece Lenses
Three recessed handpiece lenses are located at the distal end of the handpiece. Inspect the
lenses for debris prior to each treatment, as well as periodically during treatment. In both cases,
ensure that the system is turned off or in the STANDBY mode prior to inspection.
Recessed
lenses
WARNING
The system must be turned off or in the STANDBY mode prior to
inspecting the lenses. Do not inspect the lenses while the system is
in READY mode. Always wear proper eyewear while the system is
turned on. Severe eye damage can occur in the event of unintended
laser emission.
CAUTION
If the lenses are not properly maintained, debris can build up on the
lens surfaces, leading to permanent damage to the lenses and, ulti-
mately, permanent damage to the handpiece optical system.
NOTE
Debris from patients can accumulate on the lenses. Therefore, it
may be necessary to clean the lenses periodically during treatment.
Clean the handpiece lenses using a clean wood stick cotton swab and lens-cleaning grade ace-
tone, as described below.