37
GenesisPlus Operator Manual
D1151, R
EV C, 10/16
Fill the Coolant Reservoir
The system requires clean distilled water. Systems are delivered without any water in them and
must be filled with distilled water upon installation. Occasionally, distilled water may need to
be added to “top off” the system reservoir.
CAUTION
Permanent damage to internal components will occur if improper
coolant is used. One gallon of distilled water is shipped with every
system. Under no circumstances should ethylene glycol, tap water
or any liquid other than distilled water be used.
To completely fill the coolant reservoir in the system:
1. Turn off the system.
2. Remove the reservoir fill cap at the rear of the system.
3. Add water to the reservoir. Use only clean, unused distilled water provided. Carefully fill
the reservoir. Do not overfill.
4. After the reservoir is full, turn the system keyswitch to the START position and release to
the ON position.
This engages the water pump, allowing the coolant to be drawn into the cooling system.
Allow the system to run for 5 to 10 seconds.
5. Turn the keyswitch to the OFF position.
6. Top off the reservoir carefully.
7. Repeat steps 4 thru 6 until the reservoir is full.
NOTE
The coolant capacity of the system is approximately one half gal-
lon.
8. Replace the fill cap.
CAUTION
Ensure that the fill cap has a small hole to vent the system. A simi-
lar fill cap without a vent hole is used to transport the system.