36
GenesisPlus Operator Manual
D1151, R
EV C, 10/16
Inspect and Clean the Heat Exchanger Intake
The heat exchanger intake, located on the bottom of the system console, should be kept free of
dust and lint accumulation. If the heat exchanger intake becomes clogged, the system will
overheat prematurely, resulting in a temperature-related error code (i.e., E221, E224 or E229).
Inspect and clean the heat exchanger regularly using a hand mirror and long-handled bath
brush as described below.
1. Using a hand mirror, inspect the bottom of the system console for dust or lint accumula-
tion.
2. Clean the heat exchanger intake by inserting a long-handled bath brush under either side
of the system console, then pulling the brush toward your body to remove any dust or lint.
3. Repeat on the opposite side of the console to ensure that the entire heat exchanger intake is
clean.