Serial Programming Commands
10-4 PowerScan
®
7000 2D
When responding, the device echoes back the command sequence with
the status character inserted directly before each of the punctuation
marks (the period, exclamation point, comma, or semicolon) in the
command.
Examples of Query Commands
In the following examples, a bracketed notation [ ] depicts a non-dis-
playable response.
Example #1:What is the range of possible values for Codabar Coding
Enable?
Enter:cbrena*.
Response:CBRENA0-1[ACK]
This response indicates that Codabar Coding Enable (CBRENA) has a
range of values from 0 to 1 (off and on).
Example #2: What is the default value for Codabar Coding Enable?
Enter:cbrena^.
Response:CBRENA1[ACK]
This response indicates that the default setting for Codabar Coding
Enable (CBRENA) is 1, or on.
Example #3: What is the devices current setting for Codabar Coding
Enable?
Enter:cbrena?.
Response:CBRENA1[ACK]
This response indicates that the devices Codabar Coding Enable
(CBRENA) is set to 1, or on.
Example #4: What are the devices settings for all Codabar selections?
Enter:cbr?.
Response:CBRENA1[ACK],
SSX0[ACK],
CK20[ACK],
CCT1[ACK],
MIN2[ACK],
MAX60[ACK],
DFT[ACK].