EasyManuals Logo

DEWESOFT Krypton User Manual

DEWESOFT Krypton
141 pages
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Page #96 background imageLoading...
Page #96 background image
/
KRYPTON®
TECHNICAL REFERENCE MANUAL
7.4. ECAT Power Injector
The ECAT Power Injector box (article number: ECAT-POWER-INJECTOR) can be used to inject power into
the Ethernet connection between KRYPTON modules, when you can have a very long measurement
chain.
Connector Label
Connector type
Info
PWR
8-pin LEMO female
Max. 48 VDC
EtherCAT In
EtherCAT Out
8-pin LEMO
Power and Data connection to your KRYPTON module/s
use any of the ECAT to ECAT expansion cables (e.g.
L1T8m-L1T8f-10M)
GND
Banana
Ground connection
ECAT Power injector
KRYPTON® V20-1 96 / 141

Table of Contents

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the DEWESOFT Krypton and is the answer not in the manual?

DEWESOFT Krypton Specifications

General IconGeneral
SeriesKrypton
IP RatingIP67
Power Supply9 - 36 V DC
Resolution24-bit
Communication InterfaceEthernet
Input TypeAnalog, Digital, Counter, CAN

Related product manuals