EasyManua.ls Logo

Dionex ICS-1000 - Page 38

Dionex ICS-1000
210 pages
Print Icon
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
ICS-1000 Ion Chromatography System
2-20 Doc. 031879-01 3/03
device icon under the timebase, and select Properties from the Edit
menu. Select the Options tab (see Figure 2-10
).
Figure 2-10. Server Configuration Properties: Degas Mode Options
Degas Mode Options
• Always Off: The degas pump is always off.
• Always On: The degas pump is always on. This setting is for test
purposes by a Dionex Service Representative. Do not use this setting
for routine operation.
• Cycle: The degas pump cycles on and off according to the times
specified in the
Cycle On and Off fields. Cycle On specifies for how
long the degas pump runs during a cycle.
Cycle Off specifies the time
between cycles.
• Monitor: (default mode) The ICS-1000 monitors the degas pressure
reading and turns the degas pump on and off as required.

Table of Contents