Doc. 031879-01 3/03 5-1
5•Service
This chapter describes ICS-1000 service and repair procedures that the user can
perform. All procedures not included here, including electronics-related repair
procedures, must be performed by Dionex personnel. For assistance, contact
Dionex Technical Support. In the U.S., call 1-800-346-6390. Outside the U.S.,
call the nearest Dionex office.
Before replacing any part, refer to the troubleshooting information in Chapter 4
to
isolate the cause of the problem.
5.1 Diagnostic and Calibration Procedures
Diagnostic and calibration procedures are performed from the Chromeleon
Wellness Panel (see Figure 5-1
).
For an overview of the Wellness Panel features, see Section 5.1.1
. For instructions
on performing calibration and diagnostic procedures, see the following sections:
Calibrating the Conductivity Cell (Section 5.1.2)
Calibrating the Flow Rate (Section 5.1.3)
Calibrating the Vacuum Degas Assembly (Section 5.1.4)
Substituting non-Dionex parts may impair ICS-1000 performance,
thereby voiding the product warranty. Refer to the warranty statement
in the Dionex Terms and Conditions for more information.