ICS-1000 Ion Chromatography System
4-4 Doc. 031879-01 3/03
This error occurs when the degas vacuum does not reach the expected level
during calibration.
To troubleshoot:
1. Retry the calibration.
2. If calibration still fails, the vacuum degas assembly may need to be
replaced. Contact Dionex for assistance.
This error indicates an error in the ICS-1000 electronics.
To troubleshoot:
• Contact Dionex for assistance. The ICS-1000 electronics components
cannot be serviced by the user.
The leak sensor is installed in the drip tray at the bottom of the component
panel (see Figure 2-4
). If liquid accumulates in the tray, the sensor signals the
problem and the “Leak sensor wet” error message appears.
To troubleshoot:
1. Locate the source of the leak by visually inspecting the tubing,
fittings, and components on the component panel.
2. Tighten fittings or replace tubing and fittings as required. Refer to
Section 4.2
for detailed troubleshooting of various types of leaks.
3. After fixing the leak, dry the drip tray thoroughly to prevent the leak
sensor from triggering additional error messages.
DEGAS CALIBRATION FAILED
HARDWARE NOT PRESENT
LEAK SENSOR WET