4 Troubleshooting
Doc. 031879-01 3/03 4-9
This error appears normally when you click the Shutdown button on the
Chromeleon Control Panel. The message also appears anytime you turn off
the pump flow while the suppressor is on. The suppressor is automatically
turned off to prevent damage to the suppressor.
To troubleshoot:
If the pump stopped unexpectedly, see Section 4.5 for pump
troubleshooting information.
4.2 Liquid Leaks
Leaking fitting
Locate the source of the leak. Tighten or, if necessary, replace the liquid line
connection (see Section 5.3
). Refer to Installation of Dionex Liquid Line
Fittings (Document No. 031432) for tightening requirements. The manual is
on the Dionex Reference Library CD-ROM (P/N 053891).
Broken liquid line
Replace the line and fittings with the same length and internal diameter tubing
(see Section 5.3
).
Blocked or improperly installed line
Make sure the lines are not crimped or otherwise blocked. Also, if the blocked
line is a waste line, make sure it is not elevated at any point after it exits the
ICS-1000. If a line is blocked, replace it (see Section 5.3
).
Loose pump check valve
Make sure the check valves are firmly seated in the pump head. If they are
not, tighten them carefully with an open end wrench just until the leak stops.
If the leak persists, replace the check valve (see Section 5.5
).
Leaking seal wash port
Leaking from the seal wash port indicates a leaking piston seal. Replace the
seal (see Section 5.6
).
SUPPRESSOR STOPPED FOR ZERO FLOW RATE