ICS-1000 Ion Chromatography System
4-14 Doc. 031879-01 3/03
4.5 No Flow
Pump waste or priming valve open (see Figure 4-2).
Close the valves by turning the knobs clockwise until fingertight. DO NOT
OVERTIGHTEN! Overtightening may damage the valve and the pump
head.
Flow rate is set to 0
Set the flow rate from either the Chromeleon Control Panel.
Eluent valve is closed
Open the eluent valve from either the Chromeleon Control Panel.
Pump not primed
Prime the pump (see Section B.12
).
Broken pump piston
Replace the piston (see Section 5.7
).
Figure 4-2. Waste and Priming Valves