ICS-1000 Ion Chromatography System
3-10 Doc. 031879-01 3/03
5. Press Pump Off.
6. Close the waste valve. Do not overtighten. The pump is now ready for
operation.
3.7 Set System Operating Conditions
NOTE This section is an overview of the steps needed to start
up the system and begin running samples. Actual
operating parameters (flow rate, cell heater
temperature, suppressor current, etc.) depend on the
application to be run. Refer to the column manual for
the required parameters for your application.
Set or verify system operating parameters from the Chromeleon Control Panel.
Operating parameters can also be set automatically by loading a Chromeleon
sequence.
NOTE Clicking the System Startup button on the Chromeleon
Control Panel, starts the pump and suppressor. The flow
rate and suppressor current settings that were in effect
when the system was shut down are restored.
1. Verify that the pump is on and set to the correct flow rate.
2. Verify that the suppressor current is on and that the setting is correct.
3. Verify that the cell heater is set to the correct value.
4. Verify that the column heater is set to the correct value.