2 • Features
Doc. 031879-01 3/03 2-19
2.4 System Component Details
This section provides details about ICS-1000 system components, including the
vacuum degas assembly (optional), eluent valve, pump, injection valve, column
heater (optional), suppressor, and conductivity cell.
2.4.1 Vacuum Degas Assembly (Optional)
The vacuum degas assembly provides online eluent degassing at the time
and duration specified by the user. The assembly, which must be installed
in the ICS-1000 at the factory, consists of:
• A single-channel degas chamber (with degas membranes) with
internal capacity of 17 mL
• A dual-stage diaphragm vacuum pump
• A solenoid valve
• An on-board vacuum sensor
• The electronics required to operate the vacuum pump
• Tubing, fittings, and other accessories
By default, the ICS-1000 monitors the degas pressure reading and turns
the degas pump on and off as needed. A different degas operating mode
can be selected from Chromeleon. To select degas operating options from
Chromeleon, open the Server Configuration program, select the ICS-1000

Table of Contents