ICS-1000 Ion Chromatography System
4-8 Doc. 031879-01 3/03
This error appears if you turn on the suppressor and the ICS-1000 cannot
establish a connection with the suppressor.
To troubleshoot:
1. Check the suppressor cable connection (see Section 5.10
).
2. If the error persists, replace the suppressor (see Section 5.10
).
This error appears when, in order to maintain the selected current, the ICS-
1000 is required to apply a higher voltage than the suppressor can support.
To troubleshoot:
Reduce the flow rate.
Rehydrate the suppressor. Refer to the suppressor manual for instructions.
Suppressor manuals are included on the Dionex Reference Library CD-
ROM (P/N 053891), located in the ICS-1000 Ship Kit (P/N 057905).
If the error persists, replace the suppressor (see Section 5.10).
This error appears if you turn on the suppressor and the ICS-1000 cannot
establish a connection with the suppressor.
To troubleshoot:
1. Check the suppressor cable connection (see Section 5.10
).
2. If the error persists, contact Dionex for assistance. There may be a
problem in the suppressor controller electronics. The ICS-1000
electronics components cannot be serviced by the user.
SUPPRESSOR OVER VOLTAGE
SUPPRESSOR OVER POWER
SUPPRESSOR NOT CONNECTED