ICS-1000 Ion Chromatography System
2-26 Doc. 031879-01 3/03
2.4.5 Column Heater (Optional)
The column heater provides temperature
control for the separator and guard column.
The heater temperature can be set to between
30 °C and 60 °C and it must be set at least
5 °C above the ambient temperature.
Temperature is monitored via a thermistor
mounted in the heater block. Setting the
temperature to 0 °C turns off the column
heater.
If the temperature exceeds 65 °C, the column
heater is shut off and the error message
“Column heater exceeds safe temperature.”
is displayed in Chromeleon. See Section 4.1
for troubleshooting information.
The column heater can either be installed at
the factory or ordered separately (P/N
058060).
Figure 2-15. Column Heater