Operating Instructions RF-2000
29
5.2.4 Setting excitation wavelength and emission wavelength
EX(nm) EM(nm) Data
<<350>> 400
1000.00
The value for excitation wavelength is blinking in the initial state to indicate that the
excitation wavelength can be changed. Pressing the func key, emission wavelength
becomes possible to be changed and pressing shift + back keys in this state,
excitation wavelength becomes possible to be changed again. (Basically, any
blinking value shows being possible to be changed.) Enter a desired wavelength and
press Enter to change the measuring wavelength.
[Example 1]
To change the excitation wavelength from 350nm to 360nm:
(1) Ensure that the value for EX (nm) is blinking.
EX(nm) EM(nm) Data
<<350>> 400
1000.00
(2) Press the keys 3 6 0 and Enter to enter 360nm.
EX(nm) EM(nm) Data
<<360>> 400
1000.00
[Example 2]
To change the emission wavelength from 400nm to 450nm:
(1) Ensure that the value for EM (nm) is blinking by pressing the func key.
EX(nm) EM(nm) Data
360 <<400>>
1000.00
(2) Press the keys 4 5 0 and then Enter to enter 450nm.
EX(nm) EM(nm) Data
360 <<450>>
1000.00
1. When the display is not in the initial state, press the CE key to return to the
initial state.
2. Any value outside the setting range is not accepted. Setting range is 0nm and
from 200nm to 900nm and measuring range is 0nm and from 200nm to
650nm.