Operating Instructions RF-2000
28
(1) STD- func keys
zero
Auto zero key Pressing this key performs zero adjustment
of emission intensity.
zero shift
Zero shift key Zero position on the recorder or integrator
can be moved by pressing this key. It is
moved upward by pressing the
side and
downward by pressing the
side.
mark
Mark key Mark is added to the data being recorded in
the recorder and integrator by pressing this
key. Mark size: approx. 4mV (INT), approx.
4mV (REC 10mV), approx. 0.4mV (REC
1mV)
run
Run key It is a switch to start and stop a time
program.
(2) shift-func keys
shift + edit
Edit key Moves to time program edit mode.
shift + range
Range key Moves to range setting mode.
shift + sens
Sensitivity key Moves to sensitivity setting mode.
shift + scan
Scan key Wavelength scanning is performed. The
scan lamp is lit while scanning.
shift + rsp
Response key Moves to response setting mode.
(3) edit keys
0 - 9
Numerical keys Numerical values are input with these keys.
Enter
Enter key Set input value with this key.
CE
Clear key Returns the display screen to the initial
state.
Input value is cleared by pressing this key
when entering a numerical value. Display
and alarm are cleared by pressing this key
when an error is displayed. But, errors such
as equipment failure cannot be cleared with
this key.
shift + del
Delete key Deletes a line in the program displayed in
the time program.
func
Function key Advances to the next setting item in the
display screen.
AUX. FUNC setting screen is forwarded.
shift + back
Back key Returns to the previous setting item.
AUX. FUNC setting screen is scrolled
backward.
Returns to the previous setting item by
pressing this key while editing a time
program.

Table of Contents