Service Manual
Ophthalmic Surgical System
Appendix 1 - Specifications
40 - Version date: Nov. 01, 2013 - N° doc.: 30204400-B
A1.2 EVA Environmental Specifications
Electrical Specifications
Environment Specifications
Air Input
Mains supply :
100 240VAC ~50/60 Hz
Fuses:
100 240 V
T6.3AL/250V slow-blow or
T6.3AH/250V slow-blow
5 x 20 mm
Power consumption: 650 VA
Electrical Input Detachable international power cord
Equipotential grounding stud
Operation:
• An ambient temperature range of +15°C to 30°C (59°F to 86°F)
• A relative humidity range of 30% to 75% without condensation
• An atmospheric pressure range of 800hPa to 1060hPa
Storage (without
packaging):
• An ambient temperature range of +2°C to 60°C (36°F to 140°F)
• A relative humidity range of 10% to 85% without condensation
• An atmospheric pressure range of 700hPa to 1060hPa
Transport in original
packaging:
• An ambient temperature range of +2°C to 60°C (36°F to 140°F)
• A relative humidity range of 10% to 100% including condensation
• An atmospheric pressure range of 500hPa to 1060hPa
Pressure range [bar]: 4 - 10
Flow (consumption) [l]: 35 - 40
Air quality: Medical grade air