Service Manual
Ophthalmic Surgical System
Appendix 1 - Specifications
N° doc.: 30204400-B - Version date: Nov. 01, 2013 - 43
Auto Backflush (Reflux)
Manual Backflush (Reflux)
A1.6 Air module
A1.7 Vitrectomy module
A1.8 Endoillumination module
Pressure range [mmHg]: 0 to 50
Time of activation [ms]: 0 to 250
Pressure range: see Infusion /Irrigation
Pressure Range [mmHg]: 0 to 150
Pressure accuracy [mmHg]: < 3
Flow rate [ml/min]: 1200
Cut rate [cpm]: 20 to 8000
Single cut function
Number of output: 3
Maximum spectral output range [nm]: 400 - 700
Maximum intensity level (20 Ga fibre) [lm]: 40
Numerical aperture: up to NA = 0.6