Press corresponding key on the keyboard to select curve type. In which 1st order
means linear; 2nd order means quadratic regression curve; segment means
polygonal line; K-factor means k-factor input line. If select “1st order”, the
standards number should be more than 2. If select “Second order”, the standards
number should be more than 3. If select “K-factor”, then you can input the value
of k0, k1 and k2.
Note: If the standards number is not fit the curve type, the system
will automatically changed the numbers to fit the curve type.
Curve Data: Input the standards concentration.
Press <3> key on the keypad to input the Curve Data. The data list will appear
on the right side of the screen as follows. Left and right arrow key are used to
select the ID of Standards (STDs).
11/11/02 11:35 600.0 nm 0.854 ABS
Photometry
1 Parameter Setup ST
Conc
2 Sample Setup 110
3 Curve Setup 2 20
4 Print Setup 3 30
5 System Setup
6 Save Parameter
0 Forward
Input Value:_
(0 to 9999)
STD NUM: (Number of standards)
Select the numbers of standards sample. This parameter is
specifiable in range of 2 to 20.
Repeat Num: Indicates the times to repeat to measure the standard sample.
Input this value from 1-5.
3-9