8 E4D NEVO Version 5.0 10822801.C
Storage
1. Wipe water o mirror using non-woven optical wipes (recommend Kimwipes Lens Cleaning #101-7070).
2. Optional: Insert tip into a sterilization pouch. (recommend Self Seal Sterilization Pouch 5 ¼” x 10” [200/
box] #112-4854)
3. Store for later use.
Ensure the tip is dry before connecting it to the base.
Prior to scanning, visually con rm the scanning tip mirror is clean and dry.
• To clean the mirror, use a Kimwipe, 2x2 gauze, or optical cloth with a little alcohol to
gently clean the mirror.
• Dry the mirror with a dry Kimwipe, 2x2 gauze, or dry optical cloth.
Cleaning the System
For intraoral scanning systems only.
Protect the keyboard with a disposable barrier.
Cleaning Cycle: Before and after each use, clean all areas of the NEVO.
Warning: Before and after each use, follow these instructions to disinfect the NEVO. Do not
substitute any other cleaning solution or procedure. Under no circumstances should you use
any paint thinner, solvents, or harsh chemicals. Use only a non-woven sponge or pre-moistened
germicidal cloths when cleaning the NEVO.
1. Using a clean, non-woven sponge that has been saturated with a hospital grade, TB-rated germicide
or pre-moistened germicidal cloths, apply the germicide to the entire surface of the scanner base, scanner
holder, mouse, mouse pad, and any other surfaces that you touch that were not covered by a disposable
barrier. Do not spray the germicide directly on the items and do not submerge the scanner or mouse in the
germicide.
2. Follow the germicide manufacturer’s instructions.
Scanner Cradle (Laptop Systems)
Laptops come with a standalone cradle for the scanner.