Installation
20
When the analyzer’s mains power switch is set to position 2, a valve opens and the oxygen can
flow through the gas flow system. The flow rate is stabilized within a few seconds to 180 l/h and
it is displayed on the lower flow meter.
4.6 Resistance furnace-Temperature adjustment
Caution
C3.0096
Eye injury
Hot combustion tube.
Eye damage.
Avoid looking directly into the hot combustion tube. For eye protection
use the supplied protective glass.
WARNUNG
W4.0023
High temperature / Burning of the skin
Hot combustion tube
When the analyzer (HTF, resistance furnace) is switched on again after
switching off, the adjusted set point is driven immediately. Please
check this!
Depending on the material, the necessary temperature of the resistance furnace may be
different. See chapter “Applications resistance furnace”. The procedure of adjusting the
resistance furnace temperature is described below.
Fig.
8
: Resistance furnace-Temperature adjustment
Set the mains power switch to position 2.
Set the switch (14) to "Set Point" position.
Adjust 1350 on the display (15) by turning the potentiometer (13).
Set the switch (14) to "Furnace temperature" position.
The display (15) shows now the actual furnace temperature.

Table of Contents

Other manuals for ELTRA CS-2000

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the ELTRA CS-2000 and is the answer not in the manual?

ELTRA CS-2000 Specifications

General IconGeneral
BrandELTRA
ModelCS-2000
CategoryMeasuring Instruments
LanguageEnglish

Related product manuals