Maintenance
76
6.14 Gas leaks checking
Fig.
49
: Gas leaks check-Analyzer
Set the power switch (1) to pos. 2.
Close the furnace (piston up).
Press the button (43) and keep it pressed. The entire system is then checked for leaks.
After about 5 seconds of initial pressure drop, the pressure on the gauge (12) remains
constant, which means there is no leakage in the gas flow system. The leakage test is
completed. The gas system is seal.
In case there is a continuous pressure drop, then release the button (43), open the
furnace. Then press and hold the button (43) again.
If the pressure remains constant after an initial drop, then the fault is in the furnace area,
including the metal dust filter.
If the pressure still decreases, then the leakage is somewhere in the gas flow system
outside the furnace area.
Read the service manual or contact the local Eltra agent or contact Eltra GmbH directly.