Maintenance
72
6.9.2 Resistance furnace
WARNUNG
W14.0023
High temperature / Burning of the skin
Hot combustion tube
When the analyzer (HTF, resistance furnace) is switched on again after
switching off, the adjusted set point is driven immediately. Please
check this!
Fig
44
: Removing the combustion tube
CAUTION: Unplug the mains power cable before opening this device!
1. Switch off the furnace and disconnect the power plug.
2. Remove the oxygen tube (1).
3. Remove the plastic tube from the dust box (D) at the furnace outlet.
4. Dismantle the dust box by removing the screws (10) and remove the cover (11) using a
screw driver (12) if necessary.
8
3
9
6
7
1
2
D